Portable X-ray Units
Service Manual
SM-1076R2
43
5.2
ERROR CODES
Table 5-2
Error Code List
ERROR
DESCRIPTION
No Error
Code
System does not Start-up (No indication of Activity) -- Precharge Resistors.
E01
Backup Timer I2C error.
E02
Wrong Workstation configuration.
E03
No Workstation configured.
E04
Fluoro Order error.
E05
“Exposure” order is active during power-up.
E06
“Preparation” order is active during power-up.
E07
Wrong identifier for X-ray Tube 2.
E08
Wrong identifier for X-ray Tube 1.
E10
EEPROM corrupted or not initialized. Wrong data calibration.
E11
Load Capacitor error.
E12
mA is out of range during exposure.
E13
kVp is out of range during exposure.
E14
kVp ramp error.
E15
Large Filament current is out of range.
E16
Small Filament current is out of range.
E17
DC Bus is out of range.
E19
mA without exposure order.
E20
kVp without exposure order.
E23
EEPROM error.
E24
Bucky / Digital Panel error.
E25
Large Filament current demand is over the limit.
E26
Small Filament current demand is over the limit.
E27
Collimator Light current is out of range.
E28
Collimator Light current is present without order.
E29
Generator Overload error.
E30
Wrong date/time at the RTC (Real Time Clock).
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...