Portable X-ray Units
Service Manual
SM-1076R2
58
ERROR CODE :
E29
DESCRIPTION
:
Generator Overload error.
The Generator ratings are exceeded or not enough remaining Thermal Capacity of the
Generator to perform the selected exposure.
This error does not require to press the “
Reset
” push-button, the error code disappears
automatically.
POSIBLE CAUSES
The selected exposure is not possible due to the Generator thermal status.
The selected technique is beyond the Generator ratings or present conditions of the
Generator inhibit the exposure (Inverter overheated).
Parameters for next exposure may be temporally limited by the Unit.
ACTIONS
1.
Check that the remaining thermal capacity (%) is lower than the calculated for the
next exposure (thermal capacity close to 0%).
2.
Wait until the Generator cools down.
3.
If the error code does not disappear after 15 minutes, turn the Unit OFF and ON.
4.
If the equipment remains inoperative, check the hardware and replace if needed.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...