Portable X-ray Units
Service Manual
SM-1076R2
113
3.
Method-2 for disassembling the X-ray Power Unit from the Arm
:
a.
Disengage the Power / Communication Harness from the Arm
(Cable Clamps).
b.
Remove the two (2) special screws from the “U” shaped support
at both sides of the Power Unit. Take care with the washers.
c.
Separate the Power Unit with the Harness from the Arm.
d.
Remove the cover from the Connector Module of the Trolley and
disconnect the cables of the Power / Communication Harness
attached to the Power Unit.
e.
When re-assembling the Power Unit, the washers must be
installed in the same order that were removed and the two (2)
special screws must be tighten using a torque of 30 Nm.
Special Screws (2)
Power / Communication Harness
Special Screws (2)
“U” Shaped Support
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...