Portable X-ray Units
Service Manual
SM-1076R2
89
5.3.13
FILAMENTS HF BOARD
1.
Access inside the Unit by removing Upper and Lower Covers and
unscrewing the Connectors Panel (
refer to Section 5.3.1
).
2.
Remove the two Ribbon cables connected to J1 and J2 of the Filaments
HF Board (
A9500--01
).
3.
Carefully unscrew the four (4) nuts and washers that assemble the
Filaments HF Board (
A9500--01
) to the Control Driver Board
(
A3189--04
). These screws are the contacts for 0 V, +15 V, DC--BUS (+)
and DC--BUS (--). Carefully remove the board.
DC--BUS (+)
+ 15 V
0 V
DC--BUS (--)
J 1
J 2
4.
Replace the Filaments HF Board (
A9500--01
), reconnect and assemble
in reverse order.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...