Portable X-ray Units
Service Manual
SM-1076R2
12
2.2.7
DATE AND TIME (FACTORY SET AT GMT +1)
When entering the Date and Time data, keep in mind that they are
not roll over numbers.
1.
Select configuration menu C08 to enter the date/time at the RTC (Real
Time Clock).
2.
Enter sub-menu:
C8.1 for Hour Configuration (range from 0 to 23).
C8.2 for Minute Configuration (range from 0 to 59).
C8.3 for Date Configuration (range from 1 to 31).
C8.4 for Month Configuration (range from 1 to 12).
C8.5 for Year Configuration (range from 2000 to 2099).
The kVp Display shows (C8.X) and the mAs Display shows the Factory
set values.
3.
Press
“mAs Increase”
or
“mAs Decrease”
to change the values
.
4.
Press
“Reset”
push-button to save values.
5.
Press
“Collimator Light”
push-button to exit sub-menus. The kVp Display
shows (C08).
6.
Press and hold first
“Collimator Light”
push-button and then press
“kVp
Increase”
push-button, release
“kVp Increase”
and then release
“Collimator Light”
in order to exit Configuration Mode.
Note
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...