Table 3-2
Example of manual calibrated data for Filament Current in a 4 kW Unit.
mA Stations
40 kVp
80 kVp
110 kVp (or 100 kVp)
5 mA
710
700
697
6.4 mA
725
713
708
SMALL
8 mA
737
727
722
SMALL
FOCAL
SPOT
10 mA
749
738
729
SPOT
12.5 mA
763
754
--
16 mA
781
759
--
20 mA
798
764*
--
5 mA
708
697
694
6.4 mA
721
712
706
8 mA
737
728
722
10 mA
752
740
735
12.5 mA
765
756
750
16 mA
784
769
767
LARGE
FOCAL
20 mA
800
784
782
FOCAL
SPOT
25 mA
816
801
798
SPOT
32 mA
837
823
813
40 mA
853
838
829*
50 mA
872
855
--
64 mA
895
874*
--
80 mA
916
888*
--
100 mA
939
--
--
NOTES: 1) Some of the values with * should be entered manually following the numeric trend, the values with -- are not calibrated.
2) mA and kVp ranges depend on the Portable Unit model.
2.0 kW Units are calibrated at 40, 80 and 100 kVp.
4.0 kW and 8.0 kW Units are calibrated at 40, 80 and 110 kVp.
It is suggested to write down the Autocalibration or Manual
Calibration data in the table below. Keep in mind that pressing
“mA” to check a value, the data will be increased in 1 point.
Example: in order to read the Calibration value for 80 kVp, 25 mA
and Large Focal Spot, press “kVp Increase/Decrease”
push-buttons to select 80 kVp, press” mAs Increase/Decrease” to
select 25 mAs and then press “mA”. The kVp display shows 802,
the correct value is 801. To check the following mA station,
press
twice
“mAs Increase” and once “mA”.The display shows 824 while
the correct value is 823.
Note
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...