Portable X-ray Units
Service Manual
SM-1076R2
36
7.
Do not remove the Collimator Test Tool from its original position and
adjust the Light Field by moving the Collimator Unit and/or the Collimator
Light.
a.
Collimator Light must be ON during Light Field adjustment.
b.
Modify Collimator position by carefully unscrewing and screwing
the four Centering Adjustment and Security Screws (Allen) until
the Light Field coincides with the axes reference imaged (numbers
and dots) on the processed film.
Illustration 4-2
Collimator Screws
CENTERING ADJUSTMENT AND SECURITY SCREWS
c.
If adjustment is still necessary, modify position of Collimator Light
(
refer to Collimator Manual
).
8.
Repeat exposure and procedure until the result is satisfactory. The
maximum misalignment allowed is 2 % of SID (for SID 1m = 2 cm of
tolerance).
4.2
FIELD SIZE INDICATOR TEST
Before starting with the Field Size Indicator Test, the Alignment of
X-Ray Beam Test and the Alignment of Light Field with X-Ray Field
Test should be performed.
1.
Place the Portable Unit in vertical position. The Unit has to be perfectly
aligned with a Cassette Film for a Cassette of 35 x 43 cm. Check with a
level correct horizontal position of both parts.
2.
Position SID at 1 meter (use the SID Metric Tape of the Unit).
Note
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...