Portable X-ray Units
Service Manual
SM-1076R2
60
ERROR CODE :
E32
DESCRIPTION
:
I2C Bus error while trying to access the Real Time Clock (RTC).
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
It is not possible to communicate with the RTC.
ACTIONS
1.
Check on the Control Portable Board (
A3175--03
):
G
I2C Bus pull--up resistors (R66 and R67).
G
Review all the hardware connected to the I2C Bus: P89LPC922 (U56),
M41T56 (U54).
2.
Check on the Control Driver Board (
A3189--04
) the digital potentiometer AD5245
(U10) and the multiplexer MAX4586 (U12).
3.
Check the related components to all of them.
4.
Replace the defective hardware.
ERROR CODE :
E33
DESCRIPTION
:
Serial Communication error. This error only applies to Portable Units with Trolley.
The Generator has lost the serial communications with the Remote Console (3P Console
or Touch Screen Console located in the Trolley) after establishing the link. “Recoverable”
with the “
Reset
” push-button.
POSIBLE CAUSES
The Generator is not receiving the status requests from the Remote Console at the
required rate.
ACTIONS
1.
Check the cable and connections between the Generator and the Remote
Console.
2.
Check the related components to each signal of the communication cables.
3.
Replace the defective hardware.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...