Portable X-ray Units
Service Manual
SM-1076R2
23
3.2.3
CALIBRATION OF mA GAIN
mA Gain should be calibrated before performing the
autocalibration process.
1.
To enter Calibration Menu
, press and hold
“Collimator Light”
push-button and then press and release
“mAs Decrease”
push-button
(two beeps sound). (CAL PArA) appears on the displays while
“Collimator Light”
is pressed
.
Then release
“Collimator Light”
push-button. The kVp Display shows P01 and one of the Focal Spot Leds
blinks.
1
2
Press and hold Collimator Light
Press and release mAs Decrease
Release Collimator Light
3
mAs
kVp
2.
Select menu P02 by pressing the
“kVp Increase”
push-button (kVp
Display shows P02).
3.
Press
“Collimator Light”
to enter calibration of mA Gain (P02).
4.
The Unit automatically selects: 40 kVp, 20 mA, 100 ms, and Large focal
Spot. Select on the kVp Display the Filament Current value (previously
calibrated) by pressing
“mA”
or
“s”
push-buttons.
kVp
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...