Portable X-ray Units
Service Manual
SM-1076R2
135
4.
Separate the Power Unit with the “U” shaped support and the Harness
from the Arm and leave it on the table.
Fixing Screws (4)
Power / Communication Harness
“U” Shaped Support
5.
Remove the two (2) special screws from the “U” shaped support at both
sides of the Power Unit. Take care with the washers.
Special Screws (2)
6.
Replace the “U” shaped support and assemble in reverse order. When
re-assembling the Power Unit, the washers must be installed in the same
order that were removed and the two (2) special screws must be tighten
using a torque of 30 Nm.