Portable X-ray Units
Service Manual
SM-1076R2
63
ERROR CODE :
E39
DESCRIPTION
:
System error -- Power supply error.
15 V power supply fault.
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
15 V power supply goes down, noise or fault in the hardware.
ACTIONS
1.
Turn Off/ON the unit.
2.
If the error persists, replace the defective hardware.
ERROR CODE :
E40
DESCRIPTION
:
System error -- kVp unbalanced.
Anode and Cathode branches have different voltage.
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
Defective High Voltage Transformer, noise or fault in the measuring hardware.
ACTIONS
1.
Turn Off/ON the unit.
2.
If the error persists, replace the defective hardware.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...