Portable X-ray Units
Service Manual
SM-1076R2
84
5.3.9
REAL TIME CLOCK BATTERY
1.
Turn the Unit Off and Disconnect it from mains.
2.
Place the Unit on a table
.
3.
Remove the Upper Cover screws and carefully lift the Upper Cover.
4.
Take the battery BAT1 (model CR 1216, 3V) out of the Portable Control
Board (
A3175--03
) and replace it.
5.
Assemble in reverse order.