ERROR CODE :
E51
DESCRIPTION
:
Incorrect exposure time. The exposure time is above the maximum (very uncommon) or
it is so short that it is not reachable according to the present configuration (rise time and
kVp tail / high voltage capacity). “Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
The exposure time is shorter than the added of the 25% of the rise time plus the kVp tail
from the selected kVp to the 75% of that value.
ACTIONS
Change the exposure parameters increasing the exposure time.
ERROR CODE :
E60
DESCRIPTION
:
The preprogrammed exposures for Autocalibration has finished without complete the
Autocalibration process. The number of exposures to autocalibrate a mA station have
run out. This error can be only shown in Service Mode during Autocalibration process.
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
Tube instability, defective filament control hardware, noise or fault in the mA measure.
ACTIONS
If the error persists, replace the defective hardware.
ERROR CODE :
E95
DESCRIPTION
:
Power Line not good enough for operation. The Generator could not finish properly the
“Automatic Line Power Detection” procedure. No exposure will be allowed until the user
performs a successful test. “Unrecoverable”, the system remains lock.
POSIBLE CAUSES
The line impedance is too high.
ACTIONS
1.
Repeat the “Automatic Line Power Detection” procedure (
refer to the Operator
Manual
).
2.
If error persists, plug the unit to another socket line and try again.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...