Portable X-ray Units
Service Manual
SM-1076R2
51
ERROR CODE :
E10
DESCRIPTION
:
EEPROM corrupted or not initialized. Wrong data calibration
in EEPROM.
“Unrecoverable”, the system remains lock.
POSIBLE CAUSES
The EEPROM is not initialized, or the EEPROM is defective.
ACTIONS
1.
Initialize the EEPROM (
refer to Section 5.1.1
).
2.
If the error persists, replace the Portable Control Board (
A3175--03
) and initialize
the new EEPROM (
refer to Section 5.1.1
).
ERROR CODE :
E11
DESCRIPTION
:
Load Capacitor Error -- DC Bus error during Start-up.
Error while charging the load capacitors. The DC Bus voltage doesn’t reach the
correct value during Start-up. “Unrecoverable”, the system remains lock.
POSIBLE CAUSES
Defective DC Bus hardware, noise or fault at the measuring of the DC Bus voltage.
ACTIONS
1.
Turn Off/ON the unit.
2.
If the error persists, replace the defective hardware.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...