9.
Remove the Control Driver Board (
A3189--04
) as described in
Section
5.3.14
without disassembling the Filaments HF Board (
A9500--01
).
10.
From the HV Transformer, disconnect the J10 Ribbon cable, the wires
connected to TR1--2, TR1--3, TR1--1 (GND) and the three GND wires
fixed to the HV Transformer screws.
GND Wire
GND Wires
J10
TR1--3
TR1--2
TR1--1 (GND)
Control Driver Board
11.
Remove the Collimator Adaptation Ring.