Portable X-ray Units
Service Manual
SM-1076R2
71
ERROR CODE :
Err rAn
DESCRIPTION
:
Defective RAM Memory. Memory fault in U25-EEPROM Microcontroller.
“Unrecoverable”, the system restarts again after 3 seconds.
POSIBLE CAUSES
Defective RAM memory in U25-EEPROM Microcontroller (Portable Control Board
A3175--03
).
ACTIONS
1.
Turn the Unit OFF and ON.
2.
If the equipment remains inoperative, replace the Portable Control Board
(
A3175--03
)and initialize the new EEPROM (
refer to Section 5.1.1
).
ERROR CODE :
Err nrAn
DESCRIPTION
:
Defective NVRAM Memory.
Non-Volatile Memory fault in U25-EEPROM Microcontroller.
“Unrecoverable”, the system restarts again after 3 seconds.
POSIBLE CAUSES
Defective Non-Volatile RAM memory in U25-EEPROM Microcontroller (Portable Control
Board
A3175--03
).
ACTIONS
1.
Turn the Unit OFF and ON.
2.
If the equipment remains inoperative, replace the Portable Control Board
(
A3175--03
) and initialize the new EEPROM (
refer to Section 5.1.1
).
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...