Portable X-ray Units
Service Manual
SM-1076R2
90
5.3.14
CONTROL DRIVER BOARD
1.
Access inside the Unit by removing Upper and Lower Covers and
unscrewing the Connectors Panel (
refer to Section 5.3.1
).
2.
Remove the Filament HF Board (
A9500--01
) as described in
Section
5.3.13
.
3.
Disconnect from the Control Driver Board (
A3189--04
) the cables
connected to J1, J2, J3, J4, J5, GND. Then remove the four (4) nuts
attaching the cables connected to TA, TB, TC, TD bolts and take out the
cables from the bolts.
Take note of the correspondence of each cable connected to
each bolt (TA, TB, TC, TD). It is recommended to mark each
cable to avoid any mistakes when replacing the Board.
GND
J1
J5
J3
J2
J4
TD
TC
TB
TA
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...