Portable X-ray Units
Service Manual
SM-1076R2
122
5.4.9
KEYBOARD
To replace the Keyboard is not necessary to remove the X-Ray
Power Unit from the Trolley, it can be replaced in situ.
1.
Access to the Portable Control Board (
A3175--03
) as described in
Section 5.4.7 (steps 1 to 3)
.
2.
Disconnect the Keyboard cable (J13) from the Portable Control Board.
3.
Cut the tie wrap of the inner part of the Cover. It is not necessary to
disconnect the rest of cables from the Portable Control Board for
Keyboard replacement.
4.
Unscrew the Hex screws at both sides of the Portable Control Board
support and remove it.
Tie Wrap
J13
Hex screws (6)
5.
Unstick the old Keyboard from the plate support.
6.
Clean the remaining glue with alcohol from the support plate and stick the
new Keyboard.
7.
Reconnect and assemble in reverse order.
Note
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...