Portable X-ray Units
Service Manual
SM-1076R2
47
ERROR CODE :
E02
DESCRIPTION
:
Wrong Workstation configuration. One or more workstations are not properly configured,
a default value has been assigned. “Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
The workstation configuration is invalid, for example a workstation configured as
bucky/digital panel (C1.2 or C2.2) has no input selected (C1.3 or C2.3). Or the EEPROM
is not initialized, that is very uncommon.
ACTIONS
1.
Review the workstation configuration (C01 and C02 menus) and correct the
wrong value.
2.
Consider initializing the EEPROM (
refer to Section 5.1.1
). If the EEPROM is
initialized, the calibration data will be lost and the system configuration will be
overwritten with the default values.
ERROR CODE :
E03
DESCRIPTION
:
No Workstation configured. All the workstations have no Tube configured, there is no
workstation available, a default value has been assigned. “Recoverable” with the “
Reset
”
push-button.
POSIBLE CAUSES
The workstation configuration is invalid, or the EEPROM is not initialized, that is very
uncommon.
ACTIONS
1.
Review the workstation configuration (C1.1 and C2.1 submenus) and configure
at least one of them with a value different from “0”.
2.
Consider initializing the EEPROM (
refer to Section 5.1.1
). If the EEPROM is
initialized, the calibration data will be lost and the system configuration will be
overwritten with the default values.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...