Portable X-ray Units
Service Manual
SM-1076R2
11
2.2.5
MAXIMUM X-RAY TUBE POWER
The maximum X-Ray Tube Power (power curves) can be set at
70, 80, 90 or 100% of the maximum rate. This power limitation
increases the total life of the Tube.
1.
Select Configuration menu C06.
2.
Enter sub-menu 6.1. The kVp Display shows (C6.1) and the mAs Display
shows (Factory set 100).
3.
To change the Percentage Rating, press
“mAs Increase”
or
“mAs
Decrease”
push-buttons (70, 80, 90 or 100). Value factory set at 100.
4.
Press
“Reset”
push-button to save the value.
5.
Press
“Collimator Light”
push-button to exit sub-menus. The kVp Display
shows (C06).
2.2.6
MISCELLANEOUS
1.
Select configuration menu C07.
2.
Enter sub-menu C7.1 for Speaker Sound Configuration. The kVp Display
shows (C7.1) and the mAs Display shows (factory set on).
3.
Press
“mAs Increase”
or
“mAs Decrease”
to change Speaker Sound
Configuration. Press
“Reset”
push-button to save value.
4.
Press
“kVp Increase”
to enter sub-menu C7.2 for Winter / Summer
Automatic Time Update. The kVp Display shows (C7.2) and mAs shows
(Factory set yes)
.
5.
Press
“mAs Increase”
or
“mAs Decrease”
to change the Winter /
Summer Automatic Time Update
.
Press
“Reset”
push-button to save
value.
6.
Press
“kVp Increase”
to enter sub-menu C7.3 for Collimator Light ON
with PREP. The kVp Display shows (C7.3) and mAs shows (Factory set
oFF)
.
7.
Press
“mAs Increase”
or
“mAs Decrease”
to change the Collimator Light
ON with PREP
.
Press
“Reset”
push-button to save value.
8.
Press
“Collimator Light”
push-button to exit sub-menus. The kVp Display
shows (C07).
Note
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...