Portable X-ray Units
Service Manual
SM-1076R2
129
5.4.14
REPLACEMENT PARTS IN THE CENTRAL SUPPORT
5.4.14.1 ACCESS INSIDE THE CENTRAL SUPPORT
1.
Fully extend the Arm, then lay down the Trolley, with the cassettes basket
resting on the floor, and leave the Power Unit face up on a table or similar.
2.
Access inside the Unit by removing Upper and Lower Covers and
unscrewing the Protecting Plate (
refer to Section 5.4.4, steps 3. to 7.
).
3.
Unscrew the Central Support by removing the four (4) fixing screws
located at both sides and the two (2) fixing screws located at the front side
of the Central Support.
It is not necessary to remove the Collimator to access the front
side screws due to the rotatory movement of the Collimator.
Accordingly, it is not necessary to perform any Collimator
adjustments once the X-ray Power Unit is re-assembled.
Side fixing screw (4)
Central Support
Front fixing screw (2)
Note
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...