CONFIGURATION PROCEDURE
Configuration and Calibration procedures should only be
performed at field if Portable Control Board, Control Driver
Board or the High Voltage Transformer are replaced or the
EEPROM memory is re-initialized.
All procedures described in this Service Manuals are
performed from the Control Panel of the X-ray Unit.
Configuration provides the initial settings and checkout procedures that must
be carried out before starting to work with the Unit. Functional characteristics
of this Generator are defined during Configuration.
Calibration data and some configuration data are stored in the extended
memory area of the non-volatile memory U25-EEPROM Microcontroller. It is
located on the Portable Control Board (
A3175--03
).
Do not supply the main power until specifically instructed in
this document.
2.1
CONFIGURATION OF DIP SWITCHES
The Dip Switches located in the Portable Control Board
(A3175-03)
define the
Unit configuration:
Dip Switch SW1
Description
1*
ON = Service mode enabled
OFF= Service mode disabled
2
ON = Hardware errors disabled
OFF= Hardware errors enabled
3
ON = Demo mode enabled
OFF= Demo mode disabled
4
ON = Filaments disabled
OFF= Filaments enabled
* NOTE.-- Access to Service mode can be also performed using the keyboard as indicated in
Section 2.2 while dipswitch is in “OFF” position.
All Dip Switches are Factory set to OFF position for normal operation.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...