Portable X-ray Units
Service Manual
SM-1076R2
44
Table 5-2 (cont.)
Error Code List
ERROR
DESCRIPTION
E31
Wrong Time Stamp.
E32
I2C Bus error while accessing the RTC.
E33
Serial Communication error.
E34
Exposure Timer error.
E36
Presostat / Thermostat error.
E37
Tube Overload error.
E38
System error -- Chopper failure.
E39
System error -- Power supply error.
E40
System error -- kVp Unbalanced.
E41
System error -- mA Unbalanced.
E42
Corrupted Counters.
E43
Corrupted Error Log.
E44
I2C Bus error when entering the Potentiometer.
E45
I2C Bus error when entering the Multiplexor.
E46
Busy I2C Bus.
E47
APR Lite check error.
E48
Bucky / Digital Panel interface error.
E50
Exposure has been aborted by the operator or defective exposure device.
E51
Incorrect Exposure Time.
E60
The preprogrammed exposures for Autocalibration has finished without complete the Autocalibration process.
E95
Power Line not good enough for operation.
Err rAn
Defective RAM memory.
Err nrAn
Defective NVRAM memory.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...