Portable X-ray Units
Service Manual
SM-1076R2
94
5.3.15
HV TRANSFORMER
1.
Remove the special screws at both sides of the Handle. Take care with
the washers.
2.
Remove the Handle carefully (do not scratch the covers).
Special Screws
Handle
3.
If applicable remove the DAP Rails and SID Guard attached to the
Collimator (
Refer to Section 5.3.7
).
4.
Fully unscrew the four (4) Collimator Adjusting Screws and carefully put
the Collimator aside without disconnecting the cables.
5.
Access inside the Unit by removing Upper and Lower Covers and
unscrewing the Connectors Panel (
refer to Section 5.3.1
).