Portable X-ray Units
Service Manual
SM-1076R2
17
SECTION 3
CALIBRATION
Calibration procedures should only be performed at field if
Portable Control Board, Control Driver Board or the High
Voltage Transformer are replaced or the EEPROM memory is
re-initialized.
The Calibration process is as follows:
CALIBRATION PROCESS
STEP
ACTION
COMMENTS
1
Connection of the Oscilloscope for kVp and mA
measurements.
for kVp: 1 V = 30 kV
for mA: 1 V = 10 mA
(
Refer to Section 3.1
)
2
Preliminary calibration of Filament Current.
Minimum mA station in Large Focus at 40 kVp and 80 kVp.
(
Refer to Section 3.2.1).
3
Calibration of kVp Gain.
P01 (
Refer to Section 3.2.2).
4
Calibration of mA Gain.
P02 (
Refer to Section 3.2.3).
5
Autocalibration.
If the Autocalibration process is done properly, the
Calibration is finished. If not, perform the Manual
Calibration of the Filament Current (step 6).
(
Refer to Section 3.2.4).
6
Manual Calibration of Filament Current.
To be performed only when the Autocalibration did not work
properly.
All mA stations in Small Focus and Large Focus at 80 kVp,
40 kVp and 110 kVp (or 100 kVp).
(
Refer to Section 3.2.5).
The kV values to be selected in Calibration Mode are: 40, 80, 100
and 110 kVp depending on the maximum kVp of the Unit.
Note
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...