Portable X-ray Units
Service Manual
SM-1076R2
106
5.3.16.4 EMC INPUT FILTER BOARD
1.
Access inside the Central Support as described in
Section 5.3.16.1
.
2.
Disconnect the cables connected to J1, J4, J5 and J6. Cut the Tie Wraps
if needed.
3.
Loosen the two screws of the Circuit Breaker that fixes the two (2) wires
from the Power Supply by unscrewing them through the external part of
the Central Support. Disconnect both wires.
4.
Loosen without removing the two nuts which fixes the two metallic plates
in contact with the Circuit Breaker and the EMC Input Filter Board.
Disconnect the two metallic plates from the Circuit Breaker, do not
remove them from the EMC Input Filter Board.
5.
Remove the fixing nut of the Bridge Rectifier (located at the center of the
Board).
6.
Push one by one the four (4) clip pins and remove the Board.
J1
J4
J5
J6
Clip Pins (4)
Nut of Bridge Rectifier
Nuts (2) of Metallic Plates
on EMC Input Filter Board
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...