Portable X-ray Units
Service Manual
SM-1076R2
161
4.
Configure CH1 in 1V/div and CH2 in 0.1V/div. Time base in 5 ms/div;
trigger in Normal mode by positive edge of CH1.
5.
kVp and mA relation, measured in the Oscilloscope, is as follows:
G
Kvp: 1 V = 30 kV
G
mA: 1 V = 10 mA
Whenever
an
exposure
is performed
during
the
“Radiographic Parameters” test, elapse a minimum of
5 seconds between “PREP” and “EXP”, this will avoid
undesired loss of time.
6.
Turn the Unit ON.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...