5.3.16.5 INDUCTANCE CHOPPER
1.
Access inside the Central Support as described in
Section 5.3.16.1
.
2.
Cut the Tie Wrap if needed.
3.
Cut the Heat Shrinks and unsolder the two (2) wires attached keeping in
mind the correspondence of each cable.
4.
Unscrew the four fixing screws located at the side of the Central Support.
5.
Dismount the Inductance Chopper from its support.
Fixing Screws
6.
Assemble the new Inductance Chopper on its support and then in the
Central Support.
7.
Solder the two (2) wires in the respective terminals covering both cables
with the new Heat Shrinks and applying heat to reduce the Heat Shrinks.
8.
Assemble the Unit as described in
Section 5.3.16.1
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...