Portable X-ray Units
Service Manual
SM-1076R2
39
SECTION 5
TROUBLESHOOTING
The Generator contains many self-diagnostic routines which greatly facilitates
troubleshooting. The self-diagnostic functions require proper operation of the
microprocessor.
As a general rule, the first step in any troubleshooting procedure is to verify
correct power supply voltage.
If Portable Control Board, Control Driver Board or the High
Voltage Transformer are replaced or the EEPROM memory is
re-initialized, Configuration and Calibration procedures
should be performed.
5.1
GENERAL PROCEDURES
5.1.1
RE-INITIALIZATION OF EEPROM (U25)
Calibration data and some configuration data are stored in the extended
memory area of the non-volatile memory U25-EEPROM Microcontroller. It is
located on the Portable Control Board (
A3175--03
).
In case the EEPROM memory has to be re-initialized, perform the following
steps:
1.
With the Unit power “
OFF
” and the Dip-switch SW1-1 in “
OFF
” position
on the Portable Control Board, press and hold “
Reset
”, “
kVp Increase
”
and “
mAs Increase
” push-buttons at the same time.
2.
Turn the Unit ON and wait until the message “
ini EEP
” is shown in the kVp
and mAs Displays.
kVp
mAs
3.
Release the “
Reset
”, “
kVp Increase
” and “
mAs Increase
” push-buttons.
4.
Wait until the message “
ini EEP
” disappears from the kVp and mAs
Displays (it will take 30 seconds approx.).
5.
After that, the EEPROM memory has been re-initialized and it is
necessary to perform the complete Configuration (
refer to Section 2
) and
Calibration process (
refer to Section 3
).
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...