ERROR CODE :
E46
DESCRIPTION
:
Busy I2C bus. I2C Bus error, the Bus remains always busy.
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
It is not possible to communicate as the Bus is always busy.
ACTIONS
1.
Check on the Control Portable Board (
A3175--03
):
G
I2C Bus pull--up resistors (R66 and R67).
G
Review all the hardware connected to the I2C Bus: P89LPC922 (U56),
M41T56 (U54).
2.
Check on the Control Driver Board (
A3189--04
) the digital potentiometer AD5245
(U10) and the multiplexer MAX4586 (U12).
3.
Review the related components to all of them.
4.
If the error persists, replace the defective hardware.
ERROR CODE :
E47
DESCRIPTION
:
The APR Lite checksum is wrong.
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
Defective Non-Volatile RAM memory in U25-EEPROM Microcontroller (Portable Control
Board
A3175--03
) or it has not been initialized.
Defective Portable Control Board (
A3175--03
).
ACTIONS
1.
Press the “
Reset
” push-button.
2.
If the error persists, turn Off/ON the unit.
3.
If the error persists, initialize the EEPROM (
refer to Section 5.1.1
).
4.
If the error persists, replace the Portable Control Board (
A3175--03
) and initialize
the new EEPROM (
refer to Section 5.1.1
).
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...