485
Temperature Input Unit Functions
Section 11-5
Stabilization Times
required in Each Step
The following diagram shows the stabilization times (waiting times) required
when adjusting all 4 inputs.
Note
(1) The terminal block temperature stabilization time does not affect the up-
per/lower limit adjustment after changing the wiring for input 1, so the ad-
justment can be performed immediately if 30 minutes have passed since
the Temperature Input Unit's power was turned ON.
(2) To perform the adjustment procedure properly, always allow sufficient
time for the temperature to stabilize. Also allow sufficient time for devices
such as the STV, DMM, and ZERO-CON to stabilize. Refer to each de-
vice's operating manual for details.
Step
Timing chart
Stabilization time
Input Unit’s power supply ON
Wiring for input 0 adjustment
Input 0 upper/lower limit adjustment
Input 0 bias compensation
Change wiring for input 1 adjustment
Input 1 upper/lower limit adjustment
Input 1 bias compensation
Change wiring for input 2 adjustment
Input 2 upper/lower limit adjustment
Input 2 bias compensation
Change wiring for input 3 adjustment
Input 3 upper/lower limit adjustment
Input 3 bias compensation
(End of adjustment)
Wait at least 30 minutes.
(Check whether additional
time is required for devices
such as the STV, DMM,
and ZERO-CON.)
10 minutes
40 minutes
40 minutes
40 minutes
Temperature
Input Terminal’s
stabilization time
Terminal Block temperature
stabilization time 1 after wiring
Terminal Block temperature
stabilization time 2 after wiring
is changed
Terminal Block temperature
stabilization time 3 after wiring
is changed
Terminal Block temperature
stabilization time 4 after wiring
is changed
Summary of Contents for CRT1 - 10-2008
Page 2: ...CRT1 Series CompoNet Slave Units and Repeater Unit Operation Manual Revised October 2008...
Page 3: ...iv...
Page 5: ...vi...
Page 9: ...x...
Page 15: ...xvi...
Page 23: ...xxiv Conformance to EC Directives 6...
Page 51: ...28 Slave Unit Models Section 1 2...
Page 123: ...100 Connecting External I O for Slave Units Section 3 6...
Page 365: ...342 Temperature Input Units Section 7 5...
Page 381: ...358 Expansion Unit Specifications Section 8 2...
Page 418: ...395 Repeater Unit Section 10 2 Dimensions 42 3 50 3 1 27 5 95 mm...
Page 419: ...396 Repeater Unit Section 10 2...
Page 424: ...401 CX Integrator Section 11 1...
Page 575: ...552 Current Consumption Summary Appendix D 552...
Page 581: ...558 Index...
Page 583: ...560 Revision History...