System Preparation
3-9
Brightness
<1>, <2> refer to the brightness control keys with a sun marked at the top side; key
<1>, which marked a “-” on the top side, can be used to decrease the brightness;
while key<2>, which marked a “+” on the top side, can be used to increase the
brightness.
Contrast
<3>, <4> refer to the contrast control keys with a lune marked at the top side; key <3>,
which marked a “-” on the top side, can be used to decrease the contrast, while key
<4>, which marked a “+” on the top side, can be used to increase the contrast.
NOTE:
On the monitor, the brightness adjustment comes before contrast. After
readjusting the monitor’s contrast and brightness, adjust all presets and
peripheral settings.
3.6
Connecting a Probe
CAUTION:
1. Press <Freeze> to freeze an image or turn off the power of
the system before connecting/disconnecting the probe.
Otherwise, system or probe failure may occur.
2. When connecting or disconnecting a probe, place it in a
proper position, to prevent the probe from falling off or
becoming damaged.
3. Hang the probe cable to the hanger located under the control
panel to avoid excessively bending and damaging the cable.
4. Only use the probes provided by Mindray. Aftermarket
probes may result in damage or cause a fire.
NOTE:
If a probe port is not used for a long period of time, please use the dustproof
cover to protect the probe port from dust; otherwise bad contact may result.
3.6.1 Connecting a Probe
The system has four sockets for connecting probes. The leftmost one is for 4D probes,
and the other three are for the other probes.
WARNING:
The probes, cables and connectors are in proper operating
order and free from surface defects, cracks and peeling.
Otherwise, electrical shock may result.
<2>
<1>
<3>
<4>
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...