Probes and Biopsy
13-35
NGB-005
Hold the probe and the needle-guided bracket, open the grip knob of the
needle-guided bracket.
NGB-006
Metal needle-guided bracket:
(1) Screw out the nut on the guiding block to loose the guiding block and the
needle-guided bracket.
(2) Take the guiding block away in the direction of the needle tail, and then separate
the needle from the probe and the residual parts of the needle-guided bracket.
(3) Screw out the locking nut on the needle-guided bracket to loose the bracket with
the probe.
(4) Hold the probe with one hand, and then separate the needle-guided bracket from
the probe.
Plastic needle-guided bracket:
(1) Remove the guiding block slightly along the direction of the needle’s tail.
(2) Separate the residual part of the needle-guide bracket and the probe from the
needle.
(3) Remove the support of needle-guided bracket from the probe.
NGB-007
z
Metal needle-guided bracket: