Measurement
9-1
9
Measurement
You can perform measurements on a zoomed image, cine reviewing image, real-time
image, or frozen image. For measurements details, please refer to the [Advanced
Volume].
WARNING:
1. Be sure to measure areas of interest from the most
optimal image plane to avoid misdiagnosis from
inaccurate measurement values.
2. To obtain accurate Doppler flow measurement values,
make sure the transmitting beam is not perpendicular to
the flow, otherwise false readings and potential
misdiagnosis may result.
CAUTION:
1. If an image is unfrozen or the mode is changed during a
measurement, the calipers and measurement data will be
cleared from the screen. The general measurement data
will be lost (the application measurement data are stored
in the report).
2. If the system is turned off or <End Exam> is pressed
during a measurement, the data not saved will be lost.
3. In Dual-B imaging mode, the measurement results of the
merged image may be inaccurate. Therefore, the results
are provided for reference only, not for confirming a
diagnosis.
9.1 Basic
Operations
To enter/ exit measurement
Press <Caliper> to enter general measurement.
Press <Measure> key to enter application measurement.
Press <Caliper> or <Measure> again or <Esc> to exit.
Measurement result and help information
The system displays and updates measurement results in the result window.
The help information concerning measurement and calculation is displayed in the
Help Information area at the bottom of the screen.
9.2 General
Measurements
9.2.1 2D General Measurements
2D general measurements refer to general measurements on images of B, Color, Power
or iScape imaging modes. The measurements listed below can be performed:
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...