Probes and Biopsy
13-9
13.1.4 Wearing the Probe Sheath
A legally marketed probe sheath must be installed over the probe before performing
intra-cavitary and intra-operative examination. Protective barriers may be required to
minimize disease transmission. Probe sheaths are available for use with all clinical
situations where infection is a concern.
To order probe sheath, contact:
CIVCO Medical Instruments Co.
102 First Street South, Kalona, IA 52247-9589 USA Tel: 1-319-656-4447
E-mail: [email protected]
http://www.civco.com
CAUTION:
1. Be sure to cover the probe with a new (unused) probe
sheath to prevent infection during examination. If the
package of a probe sheath is open or broken, the
sterilization of the probe sheath may not be sufficient. DO
NOT use such a probe sheath.
2. The cover contains natural rubber latex and talc that can
cause allergic reactions in some individuals.
3. DO NOT use an expired probe sheath. Before using probe
sheaths, verify whether the term of validity has expired.
Method (for reference only):
1. Place an appropriate amount of gel
inside the sheath or on probe face. Poor
imaging may result if no gel is used.
2. Insert the probe into the sheath; make
sure to use proper sterile technique.
Pull cover tightly over probe face to
remove wrinkles and air bubbles,
taking care to avoid puncturing cover.
3. Secure the sheath with enclosed elastic
bands.
4. Inspect the sheath to ensure there are
no holes or tears.
Refer to the following dimension table of all probes to choose the necessary probe sheath
type.
Model Length(mm)
Width(mm)
Height(mm)
Cable Length(mm)
C5-2 112.4
76.3
25.6
2260±50mm
3C5A 114.7
75.9
29.8
2260±50mm
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...