D-2 Electrical Safety Inspection
D.2 Device Enclosure and Accessories
D.2.1 Visual
Inspection
Test Item
Acceptance Criteria
The enclosure and accessories
No physical damage to the enclosure and
accessories.
No physical damage to meters, switches,
connectors, etc.
No residue of fluid spillage (e.g., water,
coffee, chemicals, etc.).
No loose or missing parts (e.g., knobs,
dials, terminals, etc.).
D.2.2 Contextual
Inspection
Test Item
Acceptance Criteria
The enclosure and accessories
No unusual noises (e.g., a rattle inside the
case).
No unusual smells (e.g., burning or smoky
smells, particularly from ventilation holes).
No taped notes that may suggest device
deficiencies or operator concerns.
D.3 Device
Labeling
Check the labels provided by the manufacturer or the healthcare facility are present and
legible.
z
Main unit label
z
Integrated warning labels
D.4 Protective
Earth
Resistance
a. Plug the probes of the analyzer into the device’s protective earth terminal and protective
earth terminal of the AC power cord.
b. Test the earth resistance with a current of 25 A.
c. Verify the resistance is less than limits.
LIMITS
ALL COUNTRIES R = 0.2
Ω
Maximum
D.5 Earth
Leakage
Test
Run an Earth Leakage test on the device being tested before performing any other
leakage tests.
The following outlet conditions apply when performing the Earth Leakage test.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...