9-4 Measurement
9.4 Measurement
Accuracy
Table 1 Error of 2D Images
Parameter
Value Range
Error
Distance Full
screen
Within ±3%; or when the measured value is
less than 40 mm, the error is less than1.5 mm.
Area (Trace)
Full screen
Within ±7%; or when the measured value is
less than 16 cm
2
, the error is less than 1.2 cm
2
.
Area (ellipse, circle)
Full screen
Within ±7%; or when the measured value is
less than 16 cm
2
, the error is less than 1.2 cm
2
.
Angle
Full screen
Within ±3%.
Table 2 Volume Measurements
Parameter
Value
Range
Error
Volume
Full screen
Within ±10 %; or when the measured value is less than 64 cm
3
,
the error is less than 6.4 cm
3
.
Table 3 Time/Motion Measurements
Parameter
Value Range
Error
Distance
Full screen
Within ±3 %; or when the measured value is
less than 40mm, the error is less than 1.5mm.
Time
Timeline Display
Within ±2%.
Heart rate
Timeline Display
Within ±4%.
Table 4 Doppler Velocity Measurements
Parameter
Value Range
Error
Velocity (PW mode)
10-200cm/s (for V10-4,
V10-4B, L12-4, 6C2, L7-3,
7L4A, 7L5, 3C5A)
10-150cm/s (for L14-6)
10-300cm/s (for P4-2, 4CD4,
C5-2, P4-2)
When angle
≤
60º,
≤
5%.
Velocity (CW mode)
10-300cm/s (for P4-2, 2P2)
When angle
≤
60º,
≤
5%.
NOTE:
Within the selected field range, the measurement accuracy is ensured within the
range mentioned above. The accuracy specifications are performance in the
worst conditions, or based on the real test for the system, regardless of acoustic
speed error.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...