Probes and Biopsy
13-27
3. When the retaining clamp is turned to the right position, the locking nut will lock the
retaining clamp and the needle-guided bracket is then mounted to the right position.
NGB-005
(1) Inosculate the locating groove on the clamp with the two raised edges on the
probe head and aligning the locating pit of the clamp to the convex point on the
probe head.
(2) Turn the grip knob at the tail of the needle-guided bracket tightly.
NGB-006
Metal needle-guided bracket:
(1) Put on the sterile probe cover.
(2) Hold the probe by one hand, select the proper needle-guided bracket, and hold it
with the other hand. Match the groove and tab with the tab and groove of the
probe respectively. Amount the bracket onto the probe.
(3) Turn the pinch nut to secure the bracket and the probe.
(4) Select a guiding block of proper size; slightly push it into the groove on the
support.
(5) Screw the nut on the guiding block to secure the block and the needle-guided
bracket.
(6) Insert a biopsy needle with the same specification as that of the guiding block into
the hole of the guiding block.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...