5-28 Image Optimization
Display Format
Description
To set the display format of PW mode image with B mode image.
Operation
Adjust through the [Display Format] item in the soft menu or menu.
There are 4 formats to display the images: V1:1, L/R, V1:2, Full.
Impacts
The function is available in real-time imaging, freeze or cine review status.
Duplex/ Triplex
Description
This function is used to set if B image (B+Color image) and PW image are
displayed synchronously.
Operation
Turn on or off the function through the [Duplex] / [Triplex] in the soft menu
or menu.
Impacts
The function is available in real-time imaging, freeze or cine review status.
HPRF
Description
HPRF mode is used when detected velocities exceed the processing
capabilities of the currently selected PW Doppler scale or when the
selected anatomical site is too deep for the selected PW Doppler scale.
Operation
Turn on or off the function through the [HPRF] item in the soft menu or
menu.
Effects
HPRF enhances the range of detecting high-velocity flow.
Baseline
Description
Refers to the area where the velocity is zero in the spectrum.
Operation
Adjust through the [Baseline] item in the soft menu or menu.
Effects
Adjust baseline according to the actual situation to change the range of
flow velocity to optimize the image.
Positive value means to enhance the signals above the baseline, and
negative value means to enhance the signals below the baseline.
Impacts
The function is available in real-time imaging, freeze or cine review status.
Angle
Description
This function is used to adjust the angle between Doppler vector and flow
to make the velocity more accurate.
The real-time adjusting angle value is displayed on the right part of the
spectrum map.
Operation
Adjust through the [Angle] item in the soft menu or menu.
The adjustable angle range is -80~80° in increments of 1°.
Impacts
The function is available in real-time imaging, freeze or cine review status.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...