Probes and Biopsy
13-15
For details about additional infection control procedures or disinfection
information, contact
Mindray North America Customer Service [ Tel: (1-800) 288 2121, (1-201) 995
8000 ].
13.1.6 Storage and Transportation
When all examinations for the day have been completed, confirm that the probe is in good
condition. After disinfecting the probe, confirm that the probe is in good condition and
store it in a suitable place so that the next examination can be conducted smoothly.
1. To prevent the probe from being damaged, DO NOT store it in locations where it may
be exposed to:
z
Direct sunlight or X-rays
z
Sudden changes in temperature
z
Dust
z
Excessive vibration
z
Heat generators
2. Store and transport the probe under the following ambient conditions:
z
Ambient temperature: -20°C
~
55°C
z
Relative humidity: 30%
~
95% (no condensation)
z
Atmospheric pressure: 700 hPa
~
1060 hPa
Store and transport the 4CD4 probe under the following ambient conditions:
z
Ambient temperature: -10°C
~
60°C
z
Relative humidity: 30%
~
95% (no condensation)
z
Atmospheric pressure:700 hPa
~
1060 hPa
3. When the probe is sent to MINDRAY Customer Service Department or sales
representative for repair, be sure to disinfect it and keep it in the carrying case to
prevent infection.
4. Sterilize the carrying case as necessary.
13.2 Biopsy
Guide
WARNING:
1.
The person performing biopsy procedures must
understand diagnostic ultrasound thoroughly and have
been trained adequately, otherwise, side effects may be
caused to the patient.
2.
In situations listed below, the biopsy needle may fail to
penetrate the target. The incorrect biopsy may cause
various side effects in the patient.
Use a needle-guided bracket other than that provided.
Mount the needle-guided bracket incorrectly.
Use a biopsy needle that is unsuitable for the type of
biopsy being performed.
Use a biopsy needle or guiding block that is unsuitable
for the needle guide.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...