Image Optimization
5-51
z
Fan scanning
Rotate the probe once from the left to the right side (or from the right to the left) to
involve the whole region desired. See the following figure.
Parameters
description:
Type
Parameter
Description
Acquisition
Start
To begin image acquisition.
Stop
To stop image acquisition.
Parameter
adjusting
Direction
Function: to set the fast view direction in viewing mode or
in real-time displaying mode.
Selection: Up/Down, Down/ UP, Left/Right, Right/Left,
Front/Back, Back/Front.
The commonly viewed direction is Up/Down.
Display
Format
Function: to set the image display format in real-time
displaying mode (only for 4D imaging) or viewing mode.
Selection: Single, Dual, Quad
Method
Function: select the image acquisition method.
Selection: Fan, Linear.
Linear mode: during the sweep, the probe may not be
moved parallel, the speed at which you scan should be
constant.
Fan mode: in this mode, the probe must be moved to a
position where you can clearly see a middle cut of the
object you want to scan and render. Tilt the probe to about
30 degrees until the object you want to scan disappears.
Start the acquisition and tilt the probe over a distance of
around 60 degrees until the object disappears again.
During the sweep, the probe may not be moved parallel,
just tilted. The speed is about 10°/s-15°/s.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...