5-32 Image Optimization
During real-time Free Xros M mode imaging, menus for B mode, Free Xros M mode
and other modes are displayed in the soft menu at the same time, use the left/right
keys of directional soft menu controls <21> to switch the menus.
While imaging in freeze mode, menus of image optimization for B mode, Free Xros M
mode and other modes are displayed in the soft menu at the same time. Use the
left/right keys of directional soft menu controls <21> to switch the menus.
Parameters consistent with those in M mode are not to be introduced, please refer to
relevant section of the M mode, while special items of the Free Xros M mode will be
introduced in the following.
Items that appear in the menu or the soft menus are dependent upon preset, which
can be changed or set through "[Setup]
→
[Image Preset]"; please refer to "5.13
Image Preset" for details.
Display or Hide the M-mark Line
Description
There are 3 M-mark lines available, each with a symbol of "A", "B" or "C" at
the one end as identification and an arrow on the other end as a mark for
direction.
Operation
Click [Show A], [Show B], [Show C] in the soft menu or menu to display or
hide the sampling line, and the corresponding image will be displayed on
the screen once the related item is selected.
Impacts
When there is only one M-mark line on the screen, you cannot hide it.
Switching between the M-mark Lines
Description
To switch between the M-mark lines in Free Xros M mode.
Operation
Press <Set> to switch between the M-mark lines and press <Cursor> to
show the cursor.
The activated M-mark line will be green and the inactivated one will be
white.
Adjustment of the M-mark Line
Description
To adjust the position and angle of the M-mark line.
Operation
z
Position Adjustment
When the M-mark line is activated, move the trackball left and right to
adjust the position. The direction is recognized by the arrow at the end of
the line.
z
Angle Adjustment
When the M-mark line is activated, move the trackball to adjust the fulcrum
of the line, and adjust the angle via the [Angle] in the soft menu or the
multifunctional knob.
The adjusting angle range is 0-360 in increments of 1.
Display
Description
This function is to adjust the display of the images when there is more than
one M-mark line.
Operation
Adjust through the [Display] item in the soft menu or menu.
You can choose to display image of the current M-mark line or the all.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...