13-36 Probes and Biopsy
(1) Screw the nut of the guiding block and remove the guiding block slightly along the
direction of the needle’s tail.
(2) Separate the residual part of the needle-guided bracket and the probe from the
needle.
(3) Screw the pinch nut of the bracket, and remove the needle-guided bracket from
the probe.
z
Plastic needle-guided bracket:
(1) Remove the guiding block slightly along the direction of the needle’s tail.
(2) Separate the residual part of the needle-guide bracket and the probe from the
needle.
(3) Remove the support of needle-guided bracket from the probe.
NGB-010 metal/needle detachable needle-guided bracket
(1) Remove the guiding block slightly along the direction of the needle’s tail, and
separate the residual part of the needle-guide bracket and the probe from the
needle.
(2) Remove the support of needle-guided bracket from the probe.
NGB-011
Hold the probe and the needle-guided bracket, and then open the grip knob of the
needle-guided bracket.
NGB-015
(1) Pull the lock pin and open up the V-shaped cover to expose the needle.
1
2
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...