Image Optimization
5-31
5.8.1.1 Imaging
Shortcut key for Free Xros M mode
Assign a user-defined shortcut key for Free Xros M: [Setup]
→
[System Preset]
→
[Key
Config]. Please refer to "15.1.7 Key Config" for details.
Real-time
Imaging
1. In real-time B mode or M mode, adjust the probe and image to obtain the desired
plane.
2. Click [Free Xros M] in the soft menu or menu to enter Free Xros M mode, or press the
user-defined key.
3. Adjust the M-mark (single or multiple) line to obtain optimized images and necessary
information.
In Dual-split or Quad-split mode, it is the current activated window that enters the Free
Xros M Mode.
Imaging in Freeze Mode
1. Free Xros M imaging of images in cine memory
z
For B frozen images, press <M> or click [Free Xros M] in the soft menu, or press
the user-defined key directly.
z
For B+M frozen images, click [Free Xros M] in the soft menu, or press the
user-defined key directly.
z
For B+Power/Color/TVI frozen images, enter single-B cine memory first (refer to
"7.2.2 Cine Review in 2D Mode") and click [Free Xros M] in the soft menu or
press <M> on the control panel, or press the user-defined key.
z
If it is in the dual-split or quad-split scanning, it is the current activated window
that enters the Free Xros M mode, and click [Update] to switch between the
images in dual-split or quad-split B mode images.
2. Adjust the M-mark line (single or multiple) and image parameters to obtain optimized
images and necessary information.
5.8.1.2 Free
Xros
M
Image Parameters
In Free Xros M mode imaging, the image parameter area in the upper left corner of
the screen displays the real-time parameter values as follows:
Display
G 0
V 3
Parameter Free Xros M Gain Free Xros M Speed
Parameters that can be adjusted to optimize the Free Xros M mode image are
indicated in the following.
Adjustment
Items
Control Panel
Gain, TGC
Menu and Soft
Menu
Colorize, Post Process (
γ
, curve, gray rejection), Display Format,
Display, Mark Adjustment, Time Mark, Angle, Speed, Gray Map,
Colorize Map
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...