13-12 Probes and Biopsy
NOTE: Observe the graph here carefully to perform disinfection. Do not spray the strain
relief on the connector end or the connector.
Disinfecting by Immersion
1. Wear sterile gloves to prevent infection.
2. Clean the transducer before disinfecting it.MINDRAY recommends the following
solutions to disinfect the transducer.
z
Refer to the instructions provided by the chemical manufacturer concerning
concentration of the disinfectant solution, method of disinfection and dilution and
cautions during use.Do not soak the transducer connector or the cable near it into
water or any solution.
z
Soak the transducer into the disinfectant solution for the shortest time the
manufacturer recommends (for example, the shortest time recommended by the
manufacturer for soaking Cidex OPA is 12 minutes).
z
Follow local regulations when selecting and using the disinfectant.
3. Rinse the transducer with plenty of sterile water (about 2 gallons) for at least 1 minute
to remove all chemical residues on it.Or, follow the rinsing method recommended by
the disinfectant manufacturer to rinse the transducer.
4. Wipe off the water on the transducer with sterile cloth or gauze after rinsing it.Do not
dry the transducer by heating.
Connector
Strain relief
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...