5-36 Image Optimization
5.9.3 TDI Image Optimization
Parameters that can be adjusted to optimize the TDI mode image are indicated in the
following.
TDI
Operation
Items
TVI
Control Panel
Gain, Depth
Menu and Soft
Menu
Baseline, TVI IP, Acoustic Power, Line Density, B Display,
Smooth, Persistence, Focus Position, Packet Size, B/C Wide,
Dual Live, Map, Priority, WF, Frequency, Scale, Invert, Tissue
State, Line Density
TEI
Control Panel
Gain, Depth
Menu and Soft
Menu
Dual Live, TEI IP, Focus Position, Frequency, Scale, Tissue
State, Invert, WF, Persistence, Smooth, Dynamic Range, B/C
Wide, Map, Packet Size, B Display, Priority, Line Density,
Acoustic Power
TVD
Control Panel
Gain, Depth
Menu and Soft
Menu
Invert, SV, WF, Frequency, Duplex/Triplex, V Max, V Mean,
Dynamic Range, Audio, Trace Area, Colorize Map, Speed,
Acoustic Power, Trace Sensitivity, T/F Res, Gray Map, Scale,
Baseline, Colorize, Quick Angle, Angle, Trace Smooth, Time
Mark, Curve, Rejection,
γ
, Display Format, Auto Calculation,
Auto Calculation Parameter, Auto Calculation Cycle
In TVM mode, adjustable parameters are the same as those in B, M, and TVI modes;
please refer to the relevant contents for details.
Items that appear in the menu or the soft menus are dependent upon preset, which
can be changed or set through "[Setup]
→
[Image Preset]"; please refer to "5.13
Image Preset" for details.
In each TDI mode, parameters that can be adjusted are similar to those in the color flow
modes (Color, PW, and Power); please refer to relevant chapters for details. The following
is to introduce the specific items in TDI mode.
Tissue State
Description
This function is used for a fast image optimization by providing an
adjustment on 2 parameters, including scale and filter.
Operation
Adjust through the [Tissue State] item in the soft menu or menu.
There are 3 levels provided to adjust: L, M, H.
5.9.4 TDI Quantitative Analysis (QA)
CAUTION:
TDI Quantitative Analysis results are provided for reference
only, not for confirming a diagnosis. Please compare the results
with that of other machines, or make diagnosis using
none-ultrasound methods.
TDI QA is applied for TVI original data analysis, for evaluating the velocity change of the
same cardiac muscle with cardiac cycles.
Workflow of TDI QA:
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...