Image Optimization
5-21
5.6.2 Power Mode Image Parameters
In Power mode scanning, the image parameter area in the upper left corner of the
screen displays the real-time parameter values as follows:
Display
F 2.5
G 38
IP 1
WF32
PRF 0.3k
Parameter Frequency Power
Gain
Power
IP
Power Wall
Filter
Pulse Repetition
Frequency PRF
Parameters that can be adjusted to optimize the Power mode image are indicated in the
following.
Type
Parameter
Control Panel
Gain, TGC, Depth
Menu and Soft
Menu
B/C Wide, Frequency, Priority, Dynamic Range, Packet Size, Dual
Live, Power IP, Map, Acoustic Power, Focus Position, Scale, Flow
State, Wall Filter, Invert, Line Density, Smooth, Persistence, B
Display, Steer
During Power mode imaging, menus for B mode and Power mode are displayed in
the soft menu at the same time, use the left/right keys of directional soft menu controls
to switch between the menus of B mode and Power mode.
In Power mode, acoustic power is synchronous with that of B mode. Adjustment of the
depth or zoom to the B mode image will lead to corresponding changes in Power
mode image.
Items that appear in the menu or the soft menus are dependent upon preset, which
can be changed or set through "[Setup]
→
[Image Preset]"; please refer to "5.13
Image Preset" for details.
Parameters consistent with those in Color mode and B mode are not to be introduced,
please refer to relevant section of the Color mode and B mode, while special items of the
Power mode are introduced in the following.
5.6.3 Power Mode Image Optimization
Power Gain
Description
Refers to the overall sensitivity to flow signals, and this function is used to
adjust the gain in Power mode. The real-time gain value is displayed in the
image parameter area in the upper left corner of the screen.
Operation
Rotate the <Color> knob clockwise to increase the gain, and anticlockwise
to decrease.
Or adjust it in the image parameter area.
The adjusting range is 0-100.
Effects
Increasing the gain will increase the flow signal presented as well as noise,
while the signals may be missing when the gain is adjusted too low.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...