Image Optimization
5-29
Quick Angle
Description
To adjust the angle faster in increments of 60°, and the real-time value of
which is displayed on the right part of the spectrum map.
Operation
Adjust through the [Quick Angle] item in the soft menu or menu.
There are 3 angles for quickly adjustment: -60°, 0°, and 60°.
Impacts
The function is available in real-time imaging, freeze or cine review status.
Dynamic Range
Description
The dynamic range conveys the information being transformed from echo
intensity to gray scale.
Operation
Adjust through the [Dyn Ra.] item in the soft menu or menu.
The adjusting range is 24-72dB in increments of 2dB.
Effects
The more the dynamic range, the more specific the information, and the
lower the contrast with more noise.
Impacts
The function is available in real-time imaging, freeze or cine review status.
Audio
Description
This function is used to adjust the output audio in spectrum map,
Operation
Rotate the [Volume] knob in the control panel to change the audio volume.
Or press “<Fn>+ Directional key (Up/Down)” to adjust.
The adjusting range is 0-100%.
Effects
Audio helps to identify the feature and status of flow.
Impacts
The function is available in real-time imaging, freeze or cine review status.
PW Steer
Description
This function is used to adjust the scan angle of linear probes, so as to
change the angle between the transmitting beam and flow direction.
Operation
Adjust through the [PW Steer] item in the soft menu or menu.
Values of steering angles vary by probes.
Effects
This feature is used to steer the direction of the beam so as to change the
angle between the beam and flow direction with immobility of the linear
probe.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...