16-2 Batteries
“Warning! Battery is out of power! Please connect to AC power or system will be shut
down in one minute.” Connect to AC power to afford normal work.
16.3 Checking Battery Performance
The battery performance may be degraded over time, so you need to check the battery
performance periodically. The checking procedures are shown as follows:
1. Stop the ultrasound exam.
2. Connect the system to the AC power supply to charge the batteries until current
capacity is full.
3. Disconnect the system from the AC power supply to power the system by the
batteries until the system automatically shuts down.
4. The powering time of the batteries indicates the battery performance.
If the powering time of the battery is much shorter than that specified, you may replace the
battery or contact the service personnel.
NOTE:
The usage life of the battery is 300 charge / discharge cycles. If the battery is
improperly used, its life may be shortened.
16.4 Battery
Disposal
You need to dispose of the battery when it is damages or severely degraded.
NOTE:
You should observe the local regulations when disposing of the battery.
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...