Image Optimization
5-23
CW Sampling Line
Adjustment
Angle
Angle 0
°
CW Focus Depth
SVD 13.2cm
3. Set the SVD by rotating the trackball; adjust the angle and SV size according to the
actual situation.
4. Press <PW>/<CW> again or <Update> to enter PW/CW Mode and perform the
examination. You can also adjust the SV size, angle and depth in the real-time
scanning.
5. Adjust the image parameters during PW/CW mode scanning to obtain optimized
images.
6. Perform other operations (e.g. measurement and calculation) if necessary.
If you choose "Enter 1D Mode Directly" in "[Setup]
→
[Image Preset]
→
[Other]", then
the sampling line will be displayed at all times in B mode images, and pressing <PW>/
<CW> will directly enter PW/ CW mode.
5.7.2 PW/CW Mode Image Parameters
In PW/ CW mode scanning, the image parameter area in the upper left corner of the
screen displays the real-time parameter values as follows:
PW
Display
F 2.5
G 60 PRF 3.6k
WF295 SVD 8.8
SV 3.0
Parameters
Frequency Gain Pulse
Repetition
Frequency
PRF
WF
(Wall
Filter)
SV
Position
SV Size
CW Display
F 3.3
G 30
WF407
PRF 7.3k
SVD 35.0
Parameters
Frequency Gain WF
(Wall
Filter)
Pulse Repetition
Frequency PRF
SV
Position
Parameters that can be adjusted to optimize the PW/ CW mode image are indicated in the
following.
Adjustment
Items
Control Panel
Gain, TGC, Depth, Audio
Soft Menu &
Menu
PW
Invert, SV, WF, Frequency, Duplex/Triplex, V Max, V Mean, Trace
Area, Colorize Map, Speed, Acoustic Power, Trace Sensitivity, T/F
Res, Gray Map, Dynamic Range, Scale, Baseline, Colorize, Quick
Angle, Angle, Trace Smooth, Time Mark, HPRF, Curve, Gray
Rejection,
γ
, Display Format, Auto Calculation, Auto Calculation
Parameter, PW Steer
CW
WF, Invert, Colorize Map, Frequency, Speed, V Max, V Mean,
Dynamic Range, Trace Area, Trace Sensitivity, Gray Map, Scale,
Baseline, Colorize, Quick Angle, Angle, Acoustic Power, Trace
Smooth, Time Mark, T/F Res, Curve, Gray Rejection,
γ
, Display
Format, Auto Calculation, Auto Calculation Parameter
Summary of Contents for DC-T6
Page 1: ...DC T6 Diagnostic Ultrasound System Operator s Manual Basic Volume...
Page 2: ......
Page 10: ......
Page 16: ......
Page 28: ......
Page 37: ...System Overview 2 9 2 6 Introduction of Each Unit...
Page 178: ......
Page 182: ......
Page 236: ......
Page 240: ...13 4 Probes and Biopsy No Probe Model Type Illustration 19 CW2s Pencil probe...
Page 300: ......
Page 314: ......
Page 320: ......
Page 326: ......
Page 330: ...C 4 Barcode Reader...
Page 337: ...Barcode Reader C 11...
Page 342: ......
Page 347: ...P N 046 001523 01 V1 0...